![]() |
Search
NSI
|
|||||||||||||||||||||||||||||||||||||||
![]() |
|
|||||||||||||||||||||||||||||||||||||||
|
About |
1. What is the difference between gain and directivity? 2. How can I determine if a planar near-field is a good solution for my antenna tests? 3. What are the limitations of the MARS post processor? 6. What is the significance of the 3 lambda AUT/probe separation distance? 7. Compare the Direct Method Vs Comparison Method of gain calibration? 8. Do the NSI SGH NRL curves include estimates of the SGH losses? 9. When testing a CP antenna using an OEWG probe, do I need to measure both probe polarizations? 10. How critical is the phi alignment of the antenna w.r.t the probe? 11.
If my antenna is linearly polarized and I do not care about
cross-polarization, can I just take a single polarization
near-field scan, or do I still need both polarizations? Directivity is defined as the power radiated per unit solid angle in a particular direction relative to the total power radiated by the antenna. Gain is defined as the power radiated per unit solid angle in a particular direction relative to the total power accepted from the source. See the gain & directivity slides for an illustration of the concept.
2. How can I determine if a planar near-field is a good solution for my antenna tests? A few points to consider when selecting a planar near-field:
For additional information on selecting a near-field measurement system, click here.
3. What are the limitations of the MARS post processor? The MARS post processor works in conjunction with a small modification to the spherical measurement procedure. For the case of conventional spherical near-field measurements the AUT is typically installed within the range so that the majority of the current sources (i.e. the antenna) are located as close to the origin of the range axis as possible. This is done to minimize the radius of the minimum sphere which reduces the amount of data that has to be collected, and insures that the AUT is displaced as little as possible during the acquisition which tends to minimize the effects of range multipath. The MARS measurement process necessitates the displacement of the AUT from the center of rotation. This is usually accomplished by displacing the AUT in a single axis, e.g. the z-axis by a number of wavelengths (e.g. 20" - 25" @ 2.5 - 3.0 GHz). Also, the best results are achieved when the sampling increment is halved which for a fast system, roughly equates to a doubling of the acquisition time. However, the benefit of all this is that we have found that MARS is capable of improving the chamber performance by more than 10 dB!
4. What is the coordinate system reference for far-field radiation patterns obtained on a PNF range? Also, what is the ref plane when doing a back projection? The
probe reference when located at the scanner (x,y,z) = (0,0,0)
point is the coordinate reference. The reference point
on the probe is the origin of the probe coordinate system
that is used to measure or calculate the amplitude and
phase pattern of the probe. When using measured probe pattern
files, the calibration laboratory should have defined the
location of the probe coordinate system. For an OEWG this
is normally defined to be in the plane defined by the open
end of the probe with the X-Y axes parallel to the side
walls of the probe and the Z-Azis centered on the probe
parallel to the long dimension of the probe. It could be
defined differently, and in such cases the probe's phase
patterns would change but the probe corrected AUT pattern
should be independent of the definition of the probe's
coordinates for a measured probe pattern. Any change
in the X, Y or Z location of the probe origin will produce
a change in the measured probe pattern that will be equal
in magnitude and opposite in sign to the phase correction
applied to the AUT phase based on the specified location
of the probe.
5. Why is CNF (phi direction only) & SNF angular data density driven by MRE and not probe radial distance? This
is a very important concept that many customers misunderstand.
6. What is the significance of the 3 lambda AUT/probe separation distance? A NF measurement needs to be taken at a point in space where the NF probe is sampling the AUT radiating field. If the probe comes too close to the AUT it can start to detect the AUT reactive NF (fields that are coupled to the AUT and that do NOT contribute to radiation - just AUT reactance). This is typically within 1 wavelength (lambda) of the antenna. In this region the probe is not able to detect radiating fields since the reactive fields are dominant. We therefore have to ensure that the probe is outside of this region and NSI have found that 3 lambda is a safe distance. There is no maximum distance and the NF theory works even if the probe is at infinity. However, maximizing the distance leads to power loss due to space attentuation. It is accetable to reduce the distance from 3 to 2 lambda if needed, but one does run the risk of the NF probe detecing the reactive near-field, so this should be avoided unless some experimental data is available to mitigate the risk.
7. Compare the Direct Method Vs Comparison Method of gain calibration? The
connections
that should
be made for
direct gain
measurements
is to remove
the cable
from the
input to
the AUT and
the cable
from the
output connector
on the probe
and connect
these two
cables together
using the
appropriate
adapter. It
may be necessary
to use a
fixed attenuator
pad along
with the
adapter to
keep the
signal level
to the receiver
from saturating
the receiver. It
is best to
use a pad
that will
produce approximately
the same
signal level
as when the
AUT and probe
are connected
and the probe
is at the
peak amplitude
position. The
value of
the pad and
the receiver
reading should
be recorded
for entry
to the gain
calculation
menu. The
receiver
should be
left in the
same mode,
usually B/A
that it was
in when the
near‑field
measurements
were made. If
the receiver
was in the
B/R mode
for near‑field
measurements,
it should
be left in
that mode
for the
8. Do the NSI SGH NRL curves include estimates of the SGH losses? No, the NRL calculated gain curves are calculated directivity and do not include ohmic or VSWR losses. Data curves supplied with each SGH represent directivity based on the actual horn dimensions, and not gain. However, since gain assumes a perfect match we do not need be concerned about mismatch, only losses in the horn itself and this is in general in the order of 0.1 dB.
9. When testing a CP antenna using an OEWG probe, do I need to measure both probe polarizations? Yes, you need to specify in the probe setup a rectangular OEWG probe with Pol0 = 0 deg and Pol1 = 90 deg as you would when testing an LP antenna. When processing to the far-field, you need to specify the far-field principal pol sense as RHCP or LHCP and the software will combine both LP components to produce the right far-field CP sense.
10. How critical is the phi alignment of the antenna w.r.t. the probe? The phi angle is measured around the z-axis of the range measured from +X to +Y. A small phi error (dphi) has a minor effect on the principal-pol of the antenna, but a more serious impact on the cross-pol of a linear antenna. Principal-pol amplitude will decrease as the cosine of the angle whereas, cross-pol increases as the sin of the angle. The following table was generated using 20*log10(Cos(dphi)) for principal-pol and 20* log10(Sin(dphi)) for cross-pol. Note that the cross-pol level increases dramatically with dPhi (the error). This cross-pol is actually a "false" cross-pol and interferes with the ability to measure the true cross-pol of the antenna. It can be considered as a noise effect in the measurement. It will affect both cross-pol and principal measurements at low levels.
In measuring a -20dB cross-pol level or sidelobe, and assuming a 0.5 degree phi alignment error, the noise due to phi alignment error only is 41 dB. This makes the SNR between measurement and noise only 20 dB. So the measurement error will be about +/-0.9dB.
11. If my antenna is linearly polarized and I do not care about cross-polarization, can I just take a single polarization near-field scan, or do I still need both polarizations? For planar and cylindrical measurements on a linearly polarized antenna, You can often just measure the single matching polarization and get acceptable results. However for spherical near-field measurements, or if you are interested in accurate planar or cylindrical results that are off the inter-cardinal planes, you must measure both polarizations. For planar and cylindrical where the probe is polarization matched to the AUT and where the main beam is approximately normal to the plane or cylinder axis, high accuracy main component patterns can be obtained using data from the co-polarized probe only. This is true for regions near the main beam and far off axis. For an AUT with the main beam steered far off axis, the cross component probe data may be necessary for high accuracy main component patterns. For spherical measurements where the AUT is mounted with the main beam near the phi axis (polar mount), data must be obtained with both an X and a Y polarized probe. For a linearly polarized AUT mounted with its main beam normal to the phi axis (equatorial mount), the main component patterns can be obtained from data with only the co-polarized probe. The
far-field result of near-field measurements is not fully described
unless you take both polarizations. This is particularly
true for inter-cardinal cuts. Taking
only one polarization measurement will give a
reasonable replication of the far-field principal-pol
pattern under the following conditions:
12. What is the cross-pol accuracy that can be achieved using a standard OEWG probe? The OEWG has a typical cross-polarization along the two principal planes of -40 dB. If well made, it may be lower than this, but this is typical. If the OEWG is not calibrated and the analytical pattern is used in the processing, it assumes that the probe has no cross-pol. So the estimated error to signal ratio for measuring a -35 dB AUT is -5 dB. Converting this to an uncertainty in dB, gives an estimated uncertainty of 4 dB. If a lower error is required this can be achieved in two ways. If the probe is known to have a much lower cross pol level from previous measurements on this type of probe or if the manufacturer specifies a much lower level such as -60 dB, the estimated uncertainty reduces to about 1 dB. It is difficult to have high confidence in such a low cross-pol unless it is actually measured. The other way to reduce the final uncertainty is to calibrate the probe. In general, the way to reduce the AUT cross-pol uncertainty is to use a probe with the lowest cross-pol or to calibrate the probe with the best calibration accuracy. For more information, see Allen Newell's AMTA 2008 paper on cross-polarization uncertainty in near-field probe correction.
|
|||||||||||||||||||||||||||||||||||||||
| Copyright
2006 Nearfield Systems, Inc. All Rights Reserved. Please read our Legal/Privacy
Policy |
||||||||||||||||||||||||||||||||||||||||